IM 11M12A01-05E

10-9

10. Other Functions

10.3 Initialization Procedure

Follow the table below to initialize parameters. The password for initialization is 1255.

Table 10.5 Initialization Procedure

Switch operation

Display

Description

>

`

ENT

F

3

0

>

`

ENT

0

000

>

`

ENT

1

000

>

`

ENT

1

0

00

>

`

ENT

1

2

00

>

`

ENT

12

0

0

>

`

ENT

12

5

0

>

`

ENT

125

0

>

`

ENT

125

5

>

`

ENT

1255

>

`

ENT

USr Go

>

`

ENT

USr Go

>

`

ENT

F

30

T10.5E.EPS

Enter the parameter code for the item to be initialized.
The following show an example of entering "F30."
(Previous needed operations are omitted.)

Touch the [ENT] key to switch to the password entry display.

Enter the password 1255 for initialization.

After you enter the password and then touch the [ENT] key, all the digits flash.

Touch the [ENT] key again to display "USR GO."

The symbol ( ) indicates that the key is being touched.

Light characters indicate that the digits are flashing.

Touch the [ENT] key once more.
All the digits then flash for two to three seconds, and data initialization starts.

The initialization is complete, and the parameter code selection display
then appears.

WARNING

• Do not attempt to turn off the equipment power during initialization (while “USR GO”

is flashing).