5.3 Overview of display screen

(1) Measurement mode screen

On turning on the power switch, the Measurement Mode screen will appear.

The measurement screen depends on the number of components. The following screen configura-
tion shown as an example is for CO

2

, CO, O

2

(Output at channel 6).

When channel 5 or later is displayed, scroll

or

key to view another configuration which is

beyond the screen.

Fig. 5-3 Name and function of measurement mode screen

Instantaneous value and concentration value:

The concentration display of CH (component) where sampling components such as “CO

2

”,

“CO” or “O

2

are displayed in the component display, indicates current concentration values of

the measured components contained in gas that is now under measurement.

O

2

correction concentration value:

CH (components) where “cv**” is displayed as “cv CO” in the component display are calculated
from the following equation, by setting sampling components, O

2

instantaneous/concentration

values and O

2

correction reference value (see item 6.8).

Conversion output=

On:

The value of the O

2

correction referance value

(Value set by application)

Os:

Oxygen concentration (%)

Cs:

Concentration of relevant measured component

K:

The value of the fractional part is this equation.

where, K is When K

4, K=4. When K < 0, K=4.

When Cs < 0, K=0.

The correction components are CO and SO

2

only.

21 - On

21 - Os

×

Cs

K

5

6

7

8

1

2

3

4

No,

Name

Component
display

Function

Displays component of instanta-
neous value, converted instanta-
neous value, converted average
value, etc.

No,

Name

Description

Concentration
display

Displays measured value of
concentration.

Peak alarm
component
display

Displays peak alarm component.

Peak alarm
concentration
display

Displays peak alarm concentra-
tion (Upper limit value).

Range display

Displays range values.

Peak alarm
count

Displays the alarm times exceed-
ing the peak value.

Unit display

Displays unit with ppm and
vol%.

Peak alarm
unit display

Displays unit of peak alarm with
times/h.

1

2

3

4

5

6

7

8