IM 11G02M01-01E

6.6 Peak alarm setting

When the number of peaks of which CO concentration exceeds the upper limit value exceeds the
setting time, a peak alarm is outputted.

This section describes how to perform various settings of peak alarm.

(

)

ENT

End of Peak Alarm Setting

MODE

(

)

ENT

(

)

ENT

Note)

1. Press the

MODE

key in the Measurement

mode, and the User mode appears.

2. Point the cursor to “Setting of Peak Alarm”

by pressing the

or

key. Press the

ENT

key.

3. In the “Peak Alarm Setting” item selection

screen that appears, point the cursor to any
item you want by pressing the

or

key. Press the

ENT

key.

4. Then, enter numeric values and perform the

setting.

Entering or setting the numeric values
should be carried out by using the

or

key.

After setting, press the

ENT

key, and the set

values you entered are saved.

The setting is optional and is valid only
when peak alarm function is provided.

• Peak Alarm : Setting of peak alarm is performed

with ON/OFF.

• Alarm value : If peak concentrations exceed

the set alarm value, a peak counter
counts 1 time.

• Alarm Count : When a peak in excess of the

setting time occurs, a peak count
alarm output is provided.

• Hysteresis

: To prevent possible chattering

when the peak value may exceed
the set peak concentration by only
1 time, the peak count has an
allowance in the hysteresis width.

Description of setting items