2 setting of calibration gas – Yokogawa IR200 User Manual

Page 39

Advertising
background image

6 - 4

IM 11G02M01-01E

(1) Press the

MODE

key in measurement state to

display the User mode.

(2) Move the cursor to Setting about

Calibration using the

or the

key,

and then press the

ENT

key.

(3) On the Setting about Calibration screen

that appears, move the cursor to “Calibra-

tion Gas” using the

or the

key,

and then press the

ENT

key.

(4) On the “Gas Selection” screen that

appears, move the cursor to the gas to be

selected using the

or the

key,

and then press the

ENT

key.

(

)

ENT

(

)

ENT

MODE

6.2.2 Setting of calibration gas

Whether moisture is contained in zero gas and span gas used for calibration can be set in advance
as follows. Make the setting according to the sampling system created.

If sampling system controls the moisture content in the calibration gas (ZERO or SPAN) to
become the same as that in the sample gas to be fed to the analyzer using an electronic cooler,
etc., set “With” water vapor.

(

)

ENT