6 - 19

IM 11G02M01-01E

6.5 Setting of auto zero calibration

6.5.1 Auto zero calibration

Auto zero calibration is automatically carried out
when zero calibration are set.
Components for which a calibration is to be
made are determined by setting of auto calibra-
tion component in 6.2.5.

Before changing the setting of auto zero calibra-
tion, set the ON/OFF to “OFF”.

1. During measurement, press the

MODE

key

to display the User mode.

2. Point the cursor to “Setting of Auto Zero

Calibration” by pressing the

or

key. Press the

ENT

key.

3. In the “Setting of Auto Zero Calibration”

screen that appears, point the cursor to
any item you want by pressing the

or

key. Press the

ENT

key.

4. In the “Auto Zero Calibration Parameter

Setting” screen that appears, perform the
value entry or the setting. For the value
entry or setting change, use the

or

key. To change the setting, use the
key to move the cursor to the right.

After setting, press the

ENT

key, and auto

zero calibration is carried out by the
entered setting value.

(

)

ENT

MODE

(

)

ENT

ENT

End of Auto Zero Calibration Setting

Description of setting items

• Start Time : Setting at the first calibration

(day of the week, hour, minute)

• Cycle

: A period between the start time of one

calibration and another
(unit : hour/day)

• Flow Time : The time required for the calibration gas

to be replaced in the cell

• ON/OFF : Auto zero calibration ON or OFF

To close the " Auto Zero Calibration " or cancel

this mode midway, press the

ESC

key.

A previous screen will return.

To close "Auto Zero Calibration"

ESC