3 digital configuration parameters – Rice Lake Condec UMC600 User Manual

Page 17

Advertising
background image

14

UMC600 Installation Manual

3.1.3

Digital Configuration Parameters

Table 3-1 (shown below) shows an overview of the 14 configuration parameters and their description.

Table 3-2 on page 16 lists the configuration display prompts (Prompt 1), and their value selections for displayed
graduations. Prior to calibration, the UMC600 must be digitally configured, or given its set of operating parameters.
The first three parameter selections are directly related to calibration and must be set up before proceeding to the
calibration mode. These parameters include;

number of graduation

resolution

decimal point location in the weight data, all of which define the scale capacity.

Parameter

Description

Values

1

Graduations

NTEP to 10,000 (up to 80,000 available).

2

Display resolution

1, 2, 5, 10, 20, 50, 100

3

Decimal point

0.0, 0.00, 0.000, 0.0000, no decimal

5

Digital averaging

1, 2, 4, 8, 16, 32, A1 = 8-4-2; A2 = 16-8-4

6

Tare mode

ATNR, AUTO, FIXED, BOTH (inhibit with motion)

7

AZM band

Off, 0.5, 1, 3, 5, and 10 divisions
Use 0.5 for H-44, bench, counter and livestock applications; use 3 for vehicle,
axle-load, and railroad scales

8

AZM/PAZ aperture

±1.9%, 100% of capacity includes push-to-zero, H-44: 1.9%

9

Motion

Off, 1, 3, 5 divisions H-44: vehicle, axle, livestock, RR, 3.0; all other 1.0

10

Displayed units

Lb, kg, con

11

Setpoint mode

See Section 7.0 on page 55 for settings

12

Zero band

Off, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 40, 50

13

Weigh mode

Normal, truck mode with transaction cancelled/stored, fixed/auto tare

14

Serial output

Port 1, Port 2, demand, continuous, baud rate, G/T/N, or display

Table 3-1. Parameter Overview