Operation • 65

English

6. OPERATION

CAUTION!

Bodily injury

Perform a function check as described in the
section entitled "Function check" on page 76
before using the device each time.

6.1

Weighing

CAUTION!

Patients can be injured if they fall

People with restricted mobility may fall if they try
to step onto the scale.
– Support people with restricted mobility when

they step onto the scale.

– Use a suitable means of patient transport to

get those unable to stand unaided onto the
weighing platform.

– Ensure that the brakes of the means of

transport are applied during the weighing
operation.

ATTENTION!

Incorrect measurement as a result of incorrect
loading

If the scales are loaded on only one side or in
only one corner, weight will not be measured
correctly.
– Ask your patients to position themselves in

the centre of the scales.

– Position the means of patient transport

centrally on the scales.

Switching on the scale



Press the Start key.
All the elements of the display are shown briefly,
then SECA appears in the display.
The scale is operational when 0.0 appears in the
display.