English

The current content of the calibration counter
flashes in the display for a few seconds.

3. Compare the content of the calibration counter out-

put with the number given on the calibration coun-
ter sticker.

Both numbers have to match for the calibration to be
valid. If the sticker and the calibration counter do not
match, the scale must be recalibrated. Please contact
your service partner or seca service. Once the scale has
been recalibrated, a new, updated calibration counter
sticker is used to identify the status of the calibration
counter. The person authorized to perform the recali-
bration secures this sticker with an additional seal. The
calibration counter sticker can be ordered from seca
service.

11.TECHNICAL DATA

11.1 General technical data

General technical data

seca 675

Dimensions of scale

• Depth
• Width
• Height

965 mm
890 mm

65 mm

Dimensions of weighing platform

• Depth
• Width
• Height

965 mm
800 mm

65 mm

Net weight

approx. 26 kg

Temperature range

• Operation
• Storage
• Transport

+10 °C to +40 °C / +50 °F to 104 °F

-10 °C to +65 °C / +14 °F to 149 °F
-10 °C to +65 °C / +14 °F to 149 °F

Air pressure

• Operation
• Storage
• Transport

700-1060 hPa
700-1060 hPa
700-1060 hPa

Humidity

• Operation
• Storage
• Transport

30 % - 80 % no condensation

0 % - 95 % no condensation
0 % - 95 % no condensation