Troubleshooting • 77

English

9. TROUBLESHOOTING

Problem

Cause / Solution

No weight displayed
with a load on the scale.

The scale has no power supply.

- Check whether the scale is switched on.

0.0 does not appear
before weighing.

A load was already on the scale before it was
switched on.

- Remove the load from the scale.
- Switch the scale off and back on again.

One segment is illumi-
nated either continu-
ously or not at all.

The corresponding point has a fault.

- Inform the service department.

The display "StOP"
appears.

Maximum load has been exceeded.

- Remove the load from the scale.

The display "tEMP"
appears.

The ambient temperature of the scale is too high
or too low.

- Set up the scale in an ambient temperature

between +10 °C and +40 °C.

- Wait for around 15 minutes until the device

has adapted to the ambient temperature.

The display "Er:X:11"
appears.

The scale has too high a load or too high a load in
one corner.

- Take the load off the scale or distribute the

weight more evenly.

- Restart the scale.

The display "Er:X:12"
appears.

The scale has been switched on with too high a
load.

- Remove the load from the scale.
- Restart the scale.

The display "Er:X:16"
appears.

The scale was caused to oscillate, the zero point
could not be determined.

- Restart the scale.