Starting measurement – Seca 515 User Manual

Page 32

Advertising
background image

32 •

Starting measurement

1. Make an entry stating whether the patient belongs to a group of

people for whom a BIA measurement may not be performed.

yes

: The measurement is not performed. The

weight/height

tab is

active again.

no:

The procedure starts. A dialog window for positioning the

patient appears.

WARNING!
Patient hazard
Illnesses can be transmitted due to poor hygiene.
− Ensure that the patient does not have any contagious diseases.
− Ensure that the patient’s hands and feet are clean.
− Ensure that the patient does not have any open wounds on the

palms of their hands or the soles of their feet.

− Disinfect the electrode surfaces after every measurement.

WARNING!
Injury due to falling
The device is not designed to be a stand assist.
− Assist persons with limited mobility when they are sitting up e.g.

from a wheelchair.

2. Ask the patient to position themselves on the device.

3. Make sure that the patient is standing on the scale correctly:

4. Make sure that the patient is in correct contact with the hand and

foot electrodes.

Finger spacer

Rear

foot electrode

Front

foot electrode

Test point

Characteristics

Hands

• Hands must be clean
• Same pair of hand electrodes on left and right
• Select the pair of hand electrodes such that arms are

stretched out but not strained

• Finger spacers of the hand electrodes between the

middle finger and ring finger on both sides

Feet

• Stand on device with bare feet
• Feet must be clean
• Heels on the rear foot electrodes
• Balls of feet on the front foot electrodes

Position

• Upright position
• Knee slightly bent
• Do not move during the measurement