13. TROUBLESHOOTING

13.1 Power supply and display

13.2 Height and weight

Problem

Cause

Remedy

Device cannot be switched on

The device has no power supply.

Check whether power is being supplied

Power pack faulty

Replace power pack with original part

Touchscreen stays dark

Device in standby

• Touch the touchscreen display
• Press the on/off switch
• Place a load on the device

Device not switched on

Switch on the device

The device has no power supply.

Check whether power is being supplied

Touchscreen display faulty

Inform seca service

Touchscreen no longer
responds to entries or touches

Due to an implausible entry, the device
is in an undefined state.

• Switch off the device (hold the on/off

button for approx. 3 seconds)

• Switch on the device again

Display on the touchscreen
display is incomplete or
incorrect

Touchscreen display faulty

Inform seca service

Problem

Cause

Remedy

0.00 does not appear before
weighing

A load was already on the device
before switching on.

• Remove the weight from the scale
• Switch off the scale, then switch back

on again

The message

STOP appears

The maximum load has been
exceeded.

Remove the weight from the scale

The message TEMP appears

The ambient temperature is too high or
too low.

• Set up the device in an ambient

temperature between +10 °C and
+40 °C

• Wait for around 15 minutes until the

device has adapted to the ambient
temperature

The message

ER11 appears

The device has too high a load or too
high a load in one corner.

• Remove the load or distribute the load

evenly

• Restart the device

The message ER12 appears

The device has been switched on with
too high a load.

• Remove the load from the device
• Restart the device

The message

ER16 appears

The device was caused to oscillate, the
zero point could not be determined.

• Restart the device
• Restart measurement