Caution – COSA Xentaur SF6 Purity Pro User Manual

Page 11

Advertising
background image

11

2-6. AIR CAL CAUTION

If the temperature inside of the sensor has changed over ±10ºC from the last time the AIR CAL was

performed, [CAUTION/CHECK AIR CAL] will be displayed in order to prevent zero drifting.

[Measuring Gas/Base Gas] and [AIR CAL] are displayed consecutively.

If the [CAUTION/CHECK AIR CAL] is displayed, perform the AIR CAL based on the procedure below:

AIR CAL Procedure

1. Press the PUMP key and let it sample fresh air.

2. When the indication becomes stable, stop the pump with the PUMP key and press the AIR CAL

key

Caution

Before pressing the AIR CAL key, allow the instrument to sample enough fresh air (ideally for two

minutes.) Otherwise, the correct measurement cannot be acquired.

Caution

When the measurement starts, confirm if the AIR CAL value is displayed after the fresh air is

sampled enough. If the AIR CAL value is not indicated, or if [AIR CAL CAUTION] is not displayed, be

sure to perform the AIR CAL procedures correctly.