Caution – COSA Xentaur SF6 Purity Pro User Manual

Page 8

Advertising
background image

8

completed, press the PUMP key and let the sample gas draw into the

instrument. The color of the display will reverse while the pump is operating. *Note that the

indication on display under pump operation is not measuring result.

6) After the indication on the reverse display becomes stable, the pump will stop the

operation and read out the indication as a measuring result at this pump operation stop. At

this pump operation, stop under the gas flow ceasing condition. This is the measuring

result.

Caution

Press the AIR CAL key after confirming that there is enough fresh air in the chamber, or else the correct

measurement cannot be achieved.

*Note

If the [AIR] is selected as the base gas, the indication will show a zero when the AIR CAL key is

pressed. However, if the [N2] or other gases except [AIR] are selected as the base gas, the

indication will not show a zero, bur rather other values determined based on the combination of

the measuring gas and the base gas. This is caused by sampling the gas which is not on target (in

this case, the air.) Note that the instrument is not broken. Once it samples the target gas, it shows

a correct reading. Depending on the combination of the measuring gas and the base gas, the AIR

CAL value may be a negative value. For further information, refer to the measuring gas

specification.

Caution

RK60000 performs best when both the Gas Chamber and Reference Chamber are at the atmospheric

pressure (101.3 kPa.) While the measuring gas is being sampled, the pressure in the Gas Chamber

will become slightly different from the atmospheric pressure, and the analyzer will not show the correct

reading. Read the indication after the gas flow stops.

*Note

RK60000 measures the gas best when both the Gas Chamber and Reference Chamber are at

101.3 kPa. If precise measurement is required, perform the pressure correction by the following

method:

Pressure Correction = ______101.3 kPa______ X (Reading – Air Cal. Value) + Air Cal Value

Pressure at measurement [kPa]

Advertising