Type-53 [for fumigation gases – COSA Xentaur SF6 Purity Pro User Manual

Page 27

Advertising
background image

27

TYPE-53 [FOR FUMIGATION GASES]

Selectable Measuring Gas and Range

Measuring Gas

(Display Indication)

Measuring

Range

(Minimum

Digit)

AIR CAL

Value

Drift

Amount

Spec. No.

1

Methyl bromide in

AIR

(CH3Br in AIR)

0~200 mg/l

(0.2)

0.0 mg/l

1.7 mg/l

SPE- 1405

2

Phosphine in AIR

(PH3 in AIR)

0~50 mg/l

(0.1)

0.0 mg/l

0.8 mg/l

SPE- 1406

3

Methyl iodide in AIR

(CH3I in AIR)

0~200 mg/l

(0.2)

0.0 mg/l 1.7 %mg/l SPE- 1407

4

Methyl bromide in

AIR

(CH3Br in AIR)

0~5 vol%

(0.005)

0.000

vol%

0.044

vol%

SPE- 1408

5

Propylene oxide in

AIR

(PO in AIR)

0~10 vol%

(0.005)

0.000

vol%

0.038

vol%

SPE- 1448

6

Sulfuryl fluoride in

AIR

(SO2F2 in AIR)

0~200 mg/l

(0.5)

0.0 mg/l

2.9 mg/l

SPE- 1828

7

----------------

-----------------

--------------- ----------------- -----------------

8

-----------------

-----------------

--------------- ----------------- -----------------

Interferometer Type: Type120MM

Indication Accuracy: ± (Indication Value – AIR CAL Value) x 0.03 ± Drift Amount

AIR CAL Value: Concentration indicated on the display when AIR CAL key is pressed.

Drift Amount: The amount of indication drifting when temperature changes ±10ºC.

Special Caution:

1) Define Methyl bromide 2.53 vol% as 100 mg/l.

2) Define Phosphine 7.07 vol% as 100 mg/l.

3) Define Methyl iodide 1.69 vol% as 100 mg/l.

4) This instrument measures combustible gas around lower explosive limits in this

specification. Take enough sample for the measure.

5) Define Sulfuryl fluoride 2.35 vol% as 100 mg/l.