TYPE-51 [FOR SOLVENT GASES (vol%)]

Selectable Measuring Gas and Range

Measuring Gas

(Display Indication)

Measuring

Range

(Minimum

Digit)

AIR CAL

Value

Drift

Amount

Spec. No.

1

Toluene in AIR

(C7H8 in AIR)

0~2 vol%

(0.002)

0.000

vol%

0.017 vol% SPE- 1428

2

Methyl ethyl ketone

in AIR

(MEK in AIR)

0~5 vol%

(0.005)

0.000

vol%

0.027 vol% SPE- 1429

3

Ethyl acetate in AIR

(EtAc in AIR)

0~5 vol%

(0.2)

0.000

vol%

0.023 vol% SPE- 1430

4

Xylene in AIR

(C8H10 in AIR)

0~1 %

(0.002)

0.000

vol%

0.014 vol% SPE- 1431

5

Isopropyl alcohol in

AIR

(IPA in AIR)

0~2 vol%

(0.005)

0.000

vol%

0.033 vol% SPE- 1432

6

Methyl isobutyl

ketone in AIR

(MIBK in AIR)

0~1 vol%

(0.002)

0.000

vol%

0.016 vol% SPE- 1433

7

Methanol in AIR

(CH3OH in AIR)

0~6 %LEL

(0.02)

0.000

vol%

0.11 vol% SPE- 1446

8

-----------------

-----------------

--------------

----------------- -----------------

Interferometer Type: Type120MM

Indication Accuracy: ± (Indication Value – AIR CAL Value) x 0.03 ± Drift Amount

AIR CAL Value: Concentration indicated on the display when AIR CAL key is pressed.

Drift Amount: The amount of indication drifting when temperature changes ±10ºC.

Special Caution:

1) This instrument is not explosion proof.

2) This instrument measures combustible gas around lower explosive limits in this

specification. Take enough sample for the measurement.

3) Measuring gases in this specification are not extensively vaporized. The gas

concentration may not reach a higher concentration in the measuring range depending

on a condition of temperature and pressure.