487ms, 999ms, 400 series – HT instruments SPEED418 User Manual

Page 32: En - 31

Advertising
background image

400

Series

EN - 31

6. After the test is completed, if

the measured resistance
value matches the nominal
current and the set limit
contact voltage, RA<Ul/IdN
(1666Ω @ UL=50V and
IdN=30mA), the instrument
gives a double acoustic
signal and displays the
message “OK”, which

R C D

39

FRQ=50.0Hz

Ut=1.4V

VP-N=228V

VP-Pe=228V

Value of total earth resistance

Detected value for contact voltage Ut
compared to the nominal value of
the set residual current

OK

RA 30mA

50V

Func

IdN RCD UL

signals that the test has been completed successfully, and displays a screen similar to
the one reported here to the side

7.

The results displayed can be saved by pressing the SAVE key twice or the
SAVE

key and, subsequently, the ENTER key (§ 9.1)


6.4.6. Description of anomalous results

1. By using the RA mode, if the

instrument detects a contact
voltage higher than the set
limit, it displays the message
reported here to the side.
Check the efficiency of the
protective conductor and the
earthing

R C D

39

FRQ=50.0Hz

Ut=58.4V

VP-N=228V

VP-Pe=228V

Dangerous contact voltage

NOT OK

RA 30mA

50V

Func

IdN RCD UL

2. If the RCD trips in a time

exceeding the limits reported
in Tab. 5, the instrument
gives a long acoustic signal
which signals the negative
result of the test and then
displays a screen similar to
the one reported here to the
side. Check that the set type
of RCD matches the type of
RCD being tested

R C D

487ms

FRQ=50.0Hz

Ut=1.4V

VP-N=221V

VP-Pe=221V

TIME NOT OK

The tripping time is not compliant
with Tab. 5

x1 30mA

50V

Func

IdN RCD UL

3. If the RCD does not trip

within the maximum duration
of the test, the instrument
gives a long acoustic signal
which signals the negative
result of the test and then
displays a screen similar to
the one reported here to the
side. Check that the set type
of RCD matches the type of
RCD being tested

R C D

>999ms

FRQ=50.0Hz

Ut=1.4V

VP-N=221V

VP-Pe=221V

TIME NOT OK

The RCD did not trip within the
maximum duration of the test

x1 30mA

50V

Func

IdN RCD UL