Fig. 27: Instrument connection for measuring the phase sequence with two leads, phase 2 connection

9. When the instrument detects

a voltage higher than the
minimum limit value on the
test lead, the instrument
displays the screen reported
here to the side and starts
measuring the second
voltage. The instrument
gives a long acoustic signal
until input voltage is present

1 2 3

1--


Measuring…

Measuring phase 2

1T

Func

10.

Once acquisition is
completed, if the instrument
has detected a correct phase
sequence, it displays 123
and the message OK. It also
gives a double acoustic
signal

1 2 3

123


Correct phase sequence

OK

1T

Func

11.

Once acquisition is
completed, if the instrument
has detected two voltages in
phase, it displays a screen
similar to the one reported
here to the side and gives a
double acoustic signal

1 2 3

11-


Phase

conformity

OK

1T

Func

12.

The results displayed can be saved by pressing the SAVE key twice or the
SAVE

key and, subsequently, the ENTER key (§ 9.1)