Cal. report, Defaults – In-Situ Con TROLL PRO Operators Manual (purchased after 10 May 2012) User Manual

Page 31

Advertising
background image

800-446-7488

31

www.in-situ.com

Cal. Report

1. From the main screen, select Menu, Calibrate.

2. Select the Aqua TROLL 400 Instrument.

3. Select the appropriate parameter.

4. Select Cal Report to view the last time the sensor was calibrated. Select More to

view more calibration information.

Defaults

1. From the main screen, select Menu, Calibrate.

2. Select the Aqua TROLL 400 Instrument.

3. Select the appropriate parameter.

4. Select Defaults and Enter to restore the sensor to factory calibrated values.