Calibration interval, Calibrating the aqua troll 400 instrument, Calibrating the optical rdo sensor – In-Situ Con TROLL PRO Operators Manual (purchased after 10 May 2012) User Manual

Page 25: Calibration options

Advertising
background image

800-446-7488

25

www.in-situ.com

4. Select Pressure, Specific Gravity.

5. Select “Fixed” or “Auto.”

l

If you select “Fixed,” use the arrow keys to enter a desired value, and select Enter.

l

If you select Auto, the Con TROLL PRO will configure the Aqua TROLL 400 to
calculate specific gravity using the measurement from the conductivity sensor.

Calibration Interval

This function will warn you when you are overdue for a user calibration by

displaying the graph symbol

on the screen.

1. From the main screen, select Menu, Settings, Probes.

2. Select the Aqua TROLL 400 Instrument.

3. Select Options.

4. Select Cal. Interval.

5. Use the arrow keys to select the preferred interval from the on-screen options.

Calibrating the Aqua TROLL 400 Instrument

1. From the main screen, select Menu, Calibrate, then Aqua TROLL 400. You will

have the following options.

l

RDO

l

Conductivity

l

Level

l

pH/ORP 2

2. Place the calibration cap, with the vent hole, on the top of the calibration chamber.

1

Storage cap

2

Calibration cap

Figure 5.1 Storage cap and vented calibration cap shown together

Calibrating the Optical RDO Sensor

Calibration Options

From the main screen, select Menu, Calibrate, Aqua TROLL 400, RDO. You will have
the following options.

l

Saturation—Choose this option if you want to do a one-or two-point calibration.

l

Concentration—Choose this option if you want to calibrate your sensor values to
a concentration obtained using another method.

Advertising