The settings menu is accessed through the expanded
operating menu by pressing the key adjacent to the
settings icon. In the settings menu press the key below
the

wavelength icon.

This will open a number entry screen. Use the keys at the
bottom of the screen to select the digit to be adjusted.
Use the keys adjacent to the

arrow icons to increase or

decrease the wavelength to the required number. Press
the key adjacent to the

tick icon to save the changes and

return to the settings menu.

6.2.2 Settings

The settings menu enables the

wavelength, units, resolution, standard or factor to be set and is

accessed from the expanded operating menu by pressing the key adjacent to the

settings icon. Once

all of the required settings have been entered press the key adjacent to the

tick icon to save and return

to the expanded operating menu.

Fig 6.2.2.1 – Settings Menu

When setting the method parameters either the standard or the factor should be selected. The standard
should be used if the factor is not known as selecting this option will calculate the factor. If the factor is
known it is not necessary to measure a known standard’s concentration. When the standard or factor is
not selected the value should be set to 1.00.

6.2.2.1 Selecting Concentration Units

The units of concentration can be selected from a number of options: no units, %, ppm, EBC, SRM,
mEq/l, mEq, M, mM, µM, nM, U, U/l, U/ml, g/l, mg/l, µg/l, ng/l, g/dl, mg/dl, µg/dl, mg/ml, µg/ml, ng/ml,
µg/µl, ng/µl, mol/l, mmol/l.

In the settings menu press the key below the

units icon.

This opens the unit selection screen which displays all
the different units. Use the keys adjacent to the

arrow

icons to navigate around the screen to select the required
units. Once the required units have been highlighted
press the key adjacent to the

tick icon to save and return

to the settings menu. The selected unit will be displayed
in the minimal and expanded operating menu along with
absorbance and selected wavelength.

ppm

1.000

2.008

400nm

4

0

0

0

ppm

1.000

2.008

400nm

---- %

ppm

EBC SRM mEq/l

mEq M

mM

µM

nM U

U/l

U/ml g/l mg/l

µg/l ng/l g/dl

mg/dl

µg/dl mg/ml µg/ml ng/ml

µg/µl ng/µl

mol/l mmol/l.

Standard menu

Selecting
concentration units

Tick

Factor menu

Selecting a
wavelength

23

Selecting resolution