10.2

tRoUBLeSHootinG GUiDe

issue

Solution

Unable to achieve zero

Ensure that there is not a sample in the sample chamber.

absorbance or 100%

Ensure the instrument lid is closed before and during the calibration.

transmittance when

Ensure the lamp is working – if the lamp has failed replace the lamp

calibrating

(7300) or lamp module (7305).

Unable to achieve a reading Ensure the correct cuvette is being used so that light isn’t being

when measuring a sample

absorbed by the cuvette.

Ensure the sample isn’t too dense that light is not transmitted through

the sample.

Ensure the lamp is working.

Unable to print results using Ensure internal printer selected in autolog menu.
the internal printer

Ensure there is paper in the unit.

Ensure there is a result displayed on screen.


The measure sample icon

A calibration must be performed at the new wavelength. When the

disappears after changing

calibration is complete the measure sample icon will be displayed.

the wavelength