S 17 and 18 cali – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.4.08 to 4.11 User Manual

Page 139

Advertising
background image

BOP-1K 070212

4-7

11. Set the BOP to zero volts output by sending CAL:ZERO. Disconnect the +10V d-c reference

voltage from the analog I/O port external reference pin (A2A5J6 pin 11) of the BOP, then
connect 0V ±0.1mV d-c reference voltage from the external voltage source to +V_LIM_EXT
(pin 14), –V_LIM_EXT (pin 6), +I_LIM–EXT (pin 13) and –I_LIM–EXT (pin 5) referenced to
pin 12 (GND1) of the BOP analog I/O connector (A2A5J6), then send the CAL:VLIM ZERO
command. This enables the BOP to recognize that the voltage at the ±VLIM_EXT pins is 0V.

12.Set the BOP to zero amperes output by sending CAL:CLIM ZERO. This enables the BOP to

recognize that the voltage at the ±I_LIM_EXT pins is 0V. Then send CAL:ZERO command.

13.Connect +10.0V ±0.1mV d-c reference voltage from the external voltage source to

+V_LIM_EXT (pin 14), –V_LIM_EXT (pin 6), +I_LIM–EXT (pin 13) and –I_LIM–EXT (pin 5)
referenced to pin 12 (GND1) of the BOP analog I/O connector (A2A5J6), then send
CAL:VLIM MAX. This enables the BOP to recognize that the voltage at voltage at the
±VLIM_EXT pins is 10.0V. Then send CAL:ZERO command.

14.Send CAL:CLIM MAX command. This enables the BOP to recognize that the voltage at the

+I_LIM_EXT pin is 10.0V. Then send CAL:ZERO command.

15.Set the BOP to minimum (maximum negative) output by sending CAL:VLIM MIN com-

mand. This enables the BOP to recognize that the voltage at the –VLIM_EXT pin is
10.0V. Then send the CAL:ZERO command.

16.Set the BOP to minimum (maximum negative) output by sending CAL:CLIM MIN com-

mand.This enables the BOP to recognize that the voltage at the –I_LIM_EXT pin is
10.0V. Then send the CAL:ZERO command.

17.Connect a 0V ±0.1mV d-c reference to pin 8 (S_IN_SERIAL) of the PAR/SER CONTROL IN

connector (A2A5J3) referenced to COM S terminal at the rear panel (use mating connector
supplied with the unit, P/N 142-0488 to gain access to PAR/SER CONTROL IN connector
pins). Set the BOP to zero volts output by sending CAL:SER ZERO. Connect the DVM to the
BOP OUT S and OUT COM terminals and send CAL:DATA commands as needed (see PAR.
4.3a) to adjust the BOP output until the DVM reads as close to zero as possible within toler-
ance specified in Table 4-3 for VOLTAGE ZERO.

NOTE: Accuracy of the 10V d-c reference must be ±0.1mV in order for the calibrated unit to

meet published specifications.

18.Connect –10V ±0.1mV d-c reference [–6V ±0.1mV d-c reference for BOP 6-125MG] to pin 8

(S_IN_SERIAL) of the PAR/SER CONTROL IN connector (A2A5J3) referenced to COM S
terminal at the rear panel. Set the BOP to maximum positive output voltage by sending
CAL:SER MAX. Measure the voltage output using the DVM. Send CAL:DATA commands as
needed (see PAR. 4.3b) until the DVM reading is as close as possible to +10V [+6V for BOP
6-125MG].

19.Send CAL:ZERO to prepare for current calibration. After sending the command, the BOP out-

put will be set to zero volts. Refer to Figure 4-2 to connect the C terminals of the Kelvin type
sense resistor to the BOP and connect the DVM to the P (probe) terminals of the sense
resistor ((LO terminal of DVM to common P terminal). Table 4-4 provides recommended
sense resistor values for various BOP current outputs, as well as the formula for calculating
expected measured values and tolerances for any sense resistor other than those recom-
mended. Table 4-2 lists Kepco and Manufacturer part numbers for those sense resistors rec-
ommended.