PROGRAMMING/DISPLAY CHARACTERISTICS (Continued)

Trigger Port (See Table 2-3)

Shutdown (pin 2):

Operation of pin 2 is controlled by OUTP:CONT command (see Par. B.15).
Settings are STANDBY (default), HIGH, LOW, OFF. Either OFF or STANDBY is
required if using OUTP ON and OUTP OFF commands to control the output.

STANDBY (default) External Trigger Shutdown (Isolated

input. Logic 0 or short-circuit causes
the output to go to STANDBY (off,
disabled) (latched status)

Pulse width: 100

µ

S min.

Action delay: 200 mS max
Restore operation by pressing STANDBY key.
or sending OUTP ON command via remote
interface.

HIGH Isolated Input for controlling output

status, on or off. High or open (not
connected), output is on (enabled).
When 0 volts or short circuit applied
to the pin, output is off (disabled)

Pulse width: 6 mS min for any guaranteed unit
response.
Action delay: 200 mS max

LOW Isolated Input for controlling output

status, on or off. Logic 0 = output is
on (enabled); Logic 1 or open (not
connected) = output is off (disabled).

Pulse width: 6 mS min for any guaranteed unit
response.
Action delay: 200 mS max

OFF Trigger port disabled. Inputs applied

to pin 2 have no effect.

External Trigger

Input (pin 4)

Isolated input for trigger events

Pulse width: 100

µ

S min.

Action delay: 500

µ

S max

Digital control

local

Panel-mounted keypad

Direct Entry

remote

IEEE 488-2 (GPIB)

SCPI

remote

RS 232

remote

RS 485 (BITBUS)

Used for series and parallel configurations.

Display

front panel

4" backlit LCD displays all functions

remote

All parameters read back on GPIB or RS 232 buses

FUNCTION GENERATOR CHARACTERISTICS

(See PAR. 3.3.9)

Maximum number of waveforms

Local: 16

Remote: 1

Local: Saved for power-up
Remote: Lost when the unit is turned off, must
be saved on the host computer)

Maximum number of segments or basic wave-
forms per waveform

Local: 10

Remote: 126

Or maximum number of points (whichever
comes first)
• The time interval for each segment

depends on the type of waveform: period
for ramp and square waveforms, period or
fraction of period for sine and triangular
waveforms and time interval for DC levels

• Segments that cause the waveform to

exceed the maximum number of points are
not executed. For local operation they are
marked with a triple asterisk on the front
panel. For remote operation an error mes-
sage is generated.

TABLE 1-2. BOP GENERAL SPECIFICATIONS (Continued)

SPECIFICATION

RATING/DESCRIPTION

CONDITION