Advertising
background image

4-6

BOP-1K-GL 022814

3. Move the DVM to pin 15 referenced to pin 4 of the Analog I/O port. Using an adjustment

screwdriver, adjust the ZERO ADJ pot (see Figure 2-2) until the DVM reads 0V ±1mV.
Restore the DVM across OUT S and COM S as shown in Figure 4-1.

4. Set the BOP to maximum positive output voltage by sending

CAL:VOLT MAX

. Measure the

voltage output using the DVM. To adjust, send

CAL:DATA

commands as needed (see PAR.

4.3b) until the DVM reading is as close as possible above the nominal full scale value within
the tolerance specified in Table 4-3 for +FULL SCALE VOLTAGE.

5. Move the DVM to pin 15 referenced to pin 4 of the Analog I/O port. Using an adjustment

screwdriver, adjust the FULL SCALE ADJ pot (see Figure 2-2) until the DVM reads +10V
±1mV. Restore the DVM across OUT S and COM S as shown in Figure 4-1.

6. Set the BOP to maximum negative output voltage by sending

CAL:VOLT MIN

. To adjust,

send the

CAL:DATA

command as needed (see PAR. 4.3b) until the reading is as close as

possible above (absolute value) the nominal full scale value within the limits specified in
Table 4-3 for –FULL SCALE VOLTAGE.

7. Send

CAL:VPR MAX

to adjust the maximum positive voltage protection limit of the power

supply. Send the

CAL:DATA

commands as needed (see PAR. 4.3c) until the reading is as

close as possible above the nominal full scale value within the limits specified in Table 4-3 for
+FULL SCALE VPR LIMIT.

8. Send

CAL:VPR MIN

to adjust the maximum negative protection limit of the power supply.

Send the

CAL:DATA

commands as needed (see PAR. 4.3c) until the reading is as close as

possible above (absolute value) the nominal full scale value within the limits specified in
Table 4-3 for –FULL SCALE VPR LIMIT.

9. Connect a 0V ±0.1mV d-c reference to pin 11 (EXT_REF) referenced to pin 10 (SGND) of

the Analog I/O Port connector (A2A5J6). Set the BOP to zero volts output by sending

CAL:VEXT ZERO

. Send

CAL:DATA

commands as needed (see PAR. 4.3a) to adjust the

BOP output voltage until the DVM reads as close to zero as possible within tolerance speci-
fied in Table 4-3 for VOLTAGE ZERO.

NOTE: Accuracy of the 10V d-c reference must be ±0.1mV in order for the calibrated unit to
meet published specifications.

10.Replace 0V reference at pin 11 (EXT_REF) of the Analog I/O Port connector (A2A5J6) with a

+10V d-c reference. Set the BOP to maximum positive output voltage by sending

CAL:VEXT MAX

. Measure the output voltage using the DVM. To adjust, send

CAL:DATA

commands as needed (see PAR. 4.3b) until the DVM reading is as close as possible above
the nominal full scale value within the tolerance specified in Table 4-3 for +FULL SCALE
VOLTAGE.

11. Replace +10V reference at pin 11 (EXT_REF) of the Analog I/O Port connector (A2A5J6)

with a 0V ±0.1mV d-c reference. Set the BOP to zero volts output by sending

CAL:VGA ZERO

. Send

CAL:DATA

commands as needed (see PAR. 4.3a) to adjust the BOP

output until the DVM reads as close to zero as possible within tolerance specified in Table 4-
3
for VOLTAGE ZERO.

12.Replace 0V reference at pin 11 (EXT_REF) of the Analog I/O Port connector (A2A5J6) with a

+10V d-c reference. Set the BOP to maximum positive output voltage by sending

CAL:VGA MAX

. Measure the output voltage using the DVM. To adjust, send

CAL:DATA