4-7

commands as needed (see PAR. 4.3a) until the DVM reading is as close to zero as possible
within tolerance specified in Table 4-3 for VOLTAGE ZERO.

13.Set the BOP to zero volts output by sending

CAL:ZERO

. Disconnect the +10V d-c reference

voltage from the analog I/O port external reference pin (A2A5J6 pin 11) of the BOP, then con-
nect 0V ±0.1mV d-c reference voltage from the external voltage source to +V_LIM_EXT (pin
14), –V_LIM_EXT (pin 6), +I_LIM–EXT (pin 13) and –I_LIM–EXT (pin 5) referenced to pin 12
(GND1) of the BOP analog I/O connector (A2A5J6), then send the

VLIM:ZERO

command fol-

lowed by the

CAL:ZERO

command. This enables the BOP to recognize that the voltage at the

±V_LIM_EXT pin is 0V.

14.Set the BOP to zero amperes output by sending

CAL:CLIM ZERO

. Then send the

CAL:ZERO

command. This enables the BOP to recognize that the voltage at the ±I_LIM–

EXT pin is 0V.

15.Connect +10.0V ±0.1mV d-c reference voltage from the external voltage source to

+V_LIM_EXT (pin 14), –V_LIM_EXT (pin 6), +I_LIM–EXT (pin 13) and –I_LIM–EXT (pin 5)
referenced to pin 12 (GND1) of the BOP analog I/O connector (A2A5J6), then send

CAL:VLIM MAX

, then

CAL:ZERO

commands. This enables the BOP to recognize that the

voltage at the +V_LIM_EXT pin is 10.0V.

16.Send

CAL:CLIM MAX

, then

CAL:ZERO

commands. This enables the BOP to recognize that

the voltage at the +I_LIM–EXT pin is 10.0V.

17.Set the BOP to minimum (maximum negative) output by sending

CAL:VLIM MIN

, then

send the

CAL:ZERO

command. This enables the BOP to recognize that the voltage at the

–V_LIM_EXT pin is 10.0V.

18.Set the BOP to minimum (maximum negative) output by sending

CAL:CLIM MIN

, then send

the

CAL:ZERO

command. This enables the BOP to recognize that the voltage at the –I_LIM–

EXT pin is 10.0V.

19.Connect a 0V ±0.1mV d-c reference to pin 8 (S_IN_SERIAL) of the PAR/SER CONTROL IN

connector (A2A5J3) referenced to COM S terminal at the rear panel (use mating connector
supplied with the unit, P/N 142-0488 to gain access to PAR/SER CONTROL IN connector
pins). Set the BOP to zero volts output by sending

CAL:SER ZERO

. Connect the DVM to the

BOP OUT S and OUT COM terminals and send

CAL:DATA

commands as needed (see PAR.

4.3a) to adjust the BOP output until the DVM reads as close to zero as possible within toler-
ance specified in Table 4-3 for VOLTAGE ZERO.

NOTE: Accuracy of the 10V d-c reference must be ±0.1mV in order for the calibrated unit to

meet published specifications.

20.Connect a –10V ±0.1mV d-c reference to pin 8 (S_IN_SERIAL) of the PAR/SER CONTROL

IN connector (A2A5J3) referenced to COM S terminal at the rear panel. Set the BOP to max-
imum positive output voltage by sending

CAL:SER MAX

. Measure the voltage output using

the DVM. Send

CAL:DATA

commands as needed (see PAR. 4.3b) until the DVM reading is

as close as possible to +10V.

21.Send

CAL:ZERO

to prepare for current calibration. After sending the command, the BOP out-

put will be set to zero volts. Refer to Figure 4-2 to connect the C terminals of the Kelvin type
sense resistor to the BOP and connect the DVM to the P (probe) terminals of the sense

Advertising