KYORITSU 6011A User Manual

Page 4

Advertising
background image

― 2 ―

barriers on the test leads.

11

During testing it is possible that there may be a momentary degradation of

the reading due to the presence of excessive transients or discharges on
the electrical system under test. Should this be observed, the test must be
repeated to obtain a correct reading. If in doubt, contact your distributor.

12

The sliding shutter on the back of the instrument is a safety device. The

instrument should not be used if it is damaged or impaired in any way, but
returned to your distributor for attention.

13

Do not operate the function selector whilst the instrument is connected to a

circuit. If, for example, the instrument has just completed a continuity test
and an insulation test is to follow, disconnect the test leads from the circuit
before moving the selector switch.

14

Do not rotate function dial when test button is depressed. If the function

switch is inadvertently moved to a new function when the test button is
depressed or in lock-down position the test in progress will be halted. To
reset, release test button and press again to restart testing on new
function.

15

THE WIRING CHECK LED (P-E, P-N) of this instrument is to protect the user

from electrical shock resulting from incorrect connection of Line and
Neutral or Line and Earth.

When the Neutral and Earth conductors are incorrectly wired, the WIRING
CHECK LED function cannot identify the incorrect connection. Other
procedures and test must be conducted to check and confirm that the
wiring is correct prior to making measurement.
Do not use this instrument to check the correct wiring of the power supply.
Kyoritsu will not be held liable for any accident that may result from
incorrect wiring of the power supply line.