Self-diagnosis, Sensor detection, Criteria of judgment and cause – KYORITSU 6315 Quick Manual User Manual

Page 12: Causes

Advertising
background image

KEW6315

KEW6315

- 10

Self-diagnosis

If “NG” judgment is given frequently, there might be something wrong with the instrument.

instrument and refer to “

Troubleshooting

” in the instruction manual.

Sensor detection

If the detection result is NG, each sensor type will be displayed in red.

Criteria of judgment and cause

Causes

Check

Type of

current sensor

- Types of the connected current sensors are harmonized? Types of the current

sensors used for measurements

???

(cause unknown)

- Current sensors are firmly connected to the

- If any failures are in doubt:

Exchange the connections of the sensors and test again.

Connect the current sensor, for

another sensor is properly detected.

If the result "NG" is given

suspected. A defect of sensor is suspected if "NG" is given for the same

sensor connected to another

Stop using the instrument and the sensor, if any defects are in doubt, and refer

to "

Troubleshooting

"

Self-diagnosis

10 -

If “NG” judgment is given frequently, there might be something wrong with the instrument. Stop using the

in the instruction manual.

If the detection result is NG, each sensor type will be displayed in red.

Causes

Types of the connected current sensors are harmonized? Types of the current

measurements should be the same.

sensors are firmly connected to the instrument?

If any failures are in doubt:

Exchange the connections of the sensors and test again.

Connect the current sensor, for which "NG" is given, to the CH on which

another sensor is properly detected.

If the result "NG" is given for the same CH, a defect of the instrument is

suspected. A defect of sensor is suspected if "NG" is given for the same

another CH.

Stop using the instrument and the sensor, if any defects are in doubt, and refer

in the instruction manual

.