Transient, Saved data – KYORITSU 6315 Quick Manual User Manual

Page 34

Advertising
background image

KEW6315

Transient

KEW6315

- 32 -

Transient

Voltage waveforms will be monitored at approx 40ksps, gapless, to calculate and check for transient event

every 200ms. The beginning of the 200ms period where the first transient is detected is regarded as the

start of the event. If further events are not detected in the following 200ms period, the beginning of the

period is regarded as the end of the event. The detected transient is assumed to be continued between the

start to the end of event detection.

Example: Transient event detection

Saved data

When an event occurs and is detected, KEW6315 records the type of the event, start/ end time and the

values. The following data will also be recorded. The event waveform is recorded for 200ms during the

1 sec of the data refresh period.

Event waveform

Waveforms and also event data on all the chs are recorded for approx. 200ms (50Hz: 10-cycle, 60Hz:

12-cycle) at 8192 points in total. When different events occur within 1 sec, only the waveforms which

contain the highest-priority events will be recorded. However, if the same type of events occur at the

same time, the one containing the highest (deepest) values will be recorded. If the highest (deepest)

values are also the same, the one with a longer duration will be recorded. As for the channels, there is

no priority order.

Priority order: Voltage transient-> Int-> Dip-> Swell-> Inrush current

RMS variations

Voltage/ current rms value (resolution: half-cycle) variations and event data on all chs are recorded for

1 sec at data refresh.

Example: Dip detection in 800ms period:








for approx 200ms

Waveform at event start (approx. 200ms)

Waveform at event end (approx. 200ms)

r.m.s variations at the detection of

event start (approx. 1sec)

r.m.s variations at the detection of

event end (approx. 1sec)

Measured
waveform

RMS
waveform

End of transient

Threshold value

Measured
waveform

Start of transient

Duration

Advertising