KYORITSU 2417 User Manual

Page 13

Advertising
background image

̶

11

̶

NOTE

When making current measurements, keep the Transformer

Jaws fully closed. Otherwise, accurate measurements

cannot be taken.

Maximum conductor size is 40 mm in diameter.

When measuring larger current, the Transformer Jaws may

buzz. This is not a fault and does not affect the accuracy

either.

(1) Set the Range Switch to the desired "A" or " mA" position.

(2) Select the desired frequency response. WIDE or 50/60 Hz,

with the Frequency Selector Switch. (see section 6-2 for
Frequency Selector Switch)

(3) Press the Trigger to open the Transformer Jaws and clamp

onto a conductor or conductors as shown in Fig. 3. Take
the reading on the display. (This method also permits
measurements of leakage current flowing through earthing
conductors and very small current.)

Fig. 3

CAUTION

Do not exceed maximum allowable current on each current range.

(see section 3 for Specifications)