Thermo Fisher Scientific CyberScan DO 600 User Manual

Page 110

Advertising
background image

102

Figure 51: Saturation Calibration Screens

4. Rinse the probe well with de-ionized water. For best results, blot the

end of the probe to dry it. Do not touch the membrane.

5. The meter then shows saturation calibration screen. The primary

display will show the current value of measurement. Wait for the
reading to stabilize.

6. Press ENTER key to confirm the calibration. The meter

automatically calibrates to 100% air saturation and returns to the
measurement mode.

Figure 52 : Saturation Measurement Screen

Function Keys available in saturation calibration mode:

TEMP (F1)

Goes to temperature calibration

Note: If you wish to completely re-calibrate the meter, you need to clear the
previous calibration data. Press CLR-C (F2) key to clear previous calibration.
The meter shows you the confirmation screen. Press ENTER key to confirm
the deleting of previous calibration.