System setup sub-group allows you to configure general settings of the meter.
The settings are displayed in 6 pages. Press NEXT-P (F2) and PREV-P (F1) to
navigate through these pages.

Parameter

Description

Factory Default

STABLE

Indicator

ENABLE - The meter displays ‘Stable’ indicator in the
measurement screen as per the ‘STABLE CRITERIA’
defined below.

DISABLE – ‘Stable’ indicator does not appear.

ENABLE

Stability

Criteria

SLOW – The reading is stabilized slowly and exhibits
good repeatability

MEDIUM – Reading stability is averaged between
slow & fast stability

FAST – Reading is stabilized quickly at the cost of
repeatability.

(This parameter has no effect if ‘STABLE’ parameter
is disabled)

FAST

Auto Hold

ENABLE - The meter holds the reading in the
measurement screen, if the reading is ‘Stable’ for 5
seconds.

If this is enabled, ‘Response time’ appears in the
measurement screen, indicating the average response
time of the probe.

DISABLE – The reading is not held

(This parameter has no effect if ‘STABLE’ parameter
is disabled. The response time may not work if the
system time has not been set as described previously.

DISABLE

Tem. Display

from

Allows you to select temperature from pH/COND/DO
probes to display in multi measurement screen.

For PCD 650-pH/COND->DO

-

Display

setting:-

Allows you to select multi modes that you would like to
be displayed on the 1st and 2

nd

row of the

measurement screen after calibration.

1

st

Row : pH / Ion / mV / Conductivity / Salinity /

Resitivity / TDS / O

2

mg/L (ppm) / O

2

(%)

2

nd

Row : pH / Ion / mV / Conductivity / Salinity /

Resitivity / TDS / O

2

mg/L (ppm) / O

2

(%)

3

rd

Row : pH / Ion / mV / Conductivity / Salinity /

Resitivity / TDS / O

2

mg/L (ppm) / O

2

(%)

-

Note: In order to activate the RESPONSE TIME function, you have to first
activate the STABLE and AUTO HOLD functions.