12

DISABLE – Meter does not turn off automatically.

ON Time

After the last key is pressed, no. of minutes the meter
should wait before automatically shuts down the meter.
Maximum range: 30 min
(This parameter has not editable if ‘AUTO OFF’ parameter
is disabled)

10 min

Back Light

(permanently

ON)

ENABLE – Sets the back light always on.
DISABLE – Sets the backlight always off.

DISABLE

Back Light

ON with (Key

press)

ENABLE – The back light of the LCD is automatically on
when any key is pressed.
DISABLE – Does not turn on the back light automatically.

DISABLE

ON time with
(Key press)

Sets the meter to wait for specified number of minutes
before automatically turning off the back light after the last
key is pressed.
(This parameter is not editable when ‘BACK LIGHT (Key
press)’ is disabled)
(This parameter has no effect if ‘BACK LIGHT (Always)’
parameter is set to ON)

1 min

System Settings Page 4 – Wireless Serial Data Communication

Figure 5: System Settings Page 4 – Wireless Serial Data Communication


This allows you to set wireless serial data communication related parameters.

Parameter

Description

Factory Default

Print Mode

IrDA – Sets serial data communication protocol to IrDA
LED – Sets serial data communication protocol to
RS232C
MEM- Logs data to meter’s memory.

IrDA

Data Format

CyberComm – Select this format if you use
CyberComm Data Acquisition Software (DAS)

CyberComm

Note: The above settings may not work if the system time has not been
set as described previously.