Thermo Fisher Scientific CyberScan DO 600 User Manual

Page 22

Advertising
background image

14

a 5-digit password in the ‘SET PASSWORD’
parameter below
DISABLE – Disable password protection of the meter

Set Pass

Word

Specify your 5-digit password here. Use

(

Up) &

(

Down) key to select a number and then press ENTER

key to confirm and move to the next digit.
Do not set your password to ‘00000’ as this is
reserved for ‘read-only’ password.
(This parameter is not editable when ‘PASSWORD
PROTECT’ is disabled)

88888

Confirm New

Password

YES – Select this if you have made changes to the
password and you wish to confirm the changes.
NO – Select this if you wish to ignore the changes
made to the password and to store the default
password.

NO




System Settings Page 6 – Data Memory & Factory Settings

Figure 7: System Settings Page 6 – Data Memory, Display Contrast & Factory Settings

This allows you to clear the memory, display contrast and reset the meter to factory
defaults.

Clear logged Data Memory: NO

Display Contrast: 12

Factory Reset: NO

Important:

1. Please memorize the password that you have entered after enabling the

password protection. Without it, you can’t disable the password protection or
reset the meter to factory defaults. However, if the user forgets his
password, he can contact the nearest distributor or Eutech
Instruments/Oakton Instruments to request for meter password
. This
would be unique to each instrument and would be tied to the serial number of
the unit.

2. Default password ‘88888’ is valid only if it is not changed with new password.
3. You can enter ‘00000’ (read-only password) if you wish to view the setup

parameters. You are not allowed to modify any parameter when you enter
‘read-only password’.


Advertising