Setting output values between 0.0 and +1.0 – Time Electronics 1090 Portable Process Calibrator User Manual

Page 12

Advertising
background image

12

1090 Technical Manual

P a g e

| 12

Inching (Incrementing/Decrementing) - arrow keys

This function sets the step size used by the Up/Down arrow keys.

Press ‘MENU’ and release

1) Auto-off Yes

Use down arrow key to display

4) xx.x

o

, xxxu

Press ‘ENTER’ to cycle between the 3 available settings

(0.1, 1.0, or 10

o

C ; 1,10 or 100uV/uA)

Battery status

Displays battery voltage

Press ‘MENU’ and release

1) Auto-off Yes

Use down arrow key to display

5) Battery x.xxV


3.14. Setting output values between 0.0 and +1.0

It is important that a leading zero is keyed in when setting output positive values between
0.0 and +1.0

The negative sign/decimal point key has a dual function of ‘-’ and ‘.’

When entering a value, if this key is the first one pressed a negative value will be entered. If
it is not the first key pressed it will enter a decimal point.

Therefore a leading zero must be entered if a positive value of between 0.0 and 1.0 is
required, eg 0.75. However, this is not the case for settings between 0 and –1.0, where it is
allowable to enter just -.75, and a leading zero is not required.