1090 calibration procedure, 1090 c alibration p rocedure – Time Electronics 1090 Portable Process Calibrator User Manual

Page 18

Advertising
background image

18

1090 Technical Manual

P a g e

| 18

6.

1090 C alibration P rocedure

Calibration


MENU 6 ENTER, enters calibration mode

The unit prompts for a password to enter cal mode The code is 982053.


There are 8 calibration steps, as follows Pressing CLEAR at any step will skip that step, leaving
the calibration data for that step unchanged. Power must not be removed during the calibration
process, as this will cause the data check sum to be incorrect.


1) Input +30mV

Apply +30.000mV to the input terminals. PRESS ENTER.

1) Input -30mV

Apply -30.000mV to the input terminals. PRESS ENTER

2) Src +80.00mV

Measure the output voltage. Use the up/down keys to adjust for an output
of 80.000mV output. For coarse the keys immediately below the arrow
keys can be used : 7/8 adjust down and up in 10 times larger units, 4/5,
100 times

When the unit is first calibrated after manufacture, the cal factor may be a
long way out. To avoid lots of key pressing, pressing SHIFT will set it to a
value close to the nominal calibration value.

3) Src +8.0mV

As above, adjust for +8.000mV

4) Src - 80.0mV

As above, adjust for -80.000mV

5) Src -8.0mV

As above, adjust for -8.000mV

6) Temp

Display current calibrated temperature. If incorrect, press ENTER and
enter the current temperature at the CJ sense point in degrees C.

7) Pt100

Only perform this task if the mA/pt100 board is fitted.

Connect a 300.0ohm precision resistor across the output terminals

Press ENTER.



the unit will then restart.