1090 Technical Manual

P a g e

| 19

If the PRT/Current board is fitted and requires calibration, this must be done after the
voltage steps 1-6 above have been performed P50 calibrates the current measure
mode, and P51 current source mode. These should be adjusted at full-scale (30mA
measure, 80mA source).

If the PRT/Current board is recalibrated, you must repeat step 8 above afterwards
(use CLEAR to skip past the other steps).

Notes on the PRT/Current option board.

The PRT/Current board has calibration adjustments on-board for current source and
measure and can be supplied ready-calibrated for retrofitting in the field.

When a 1090 unit is initially factory calibrated, it is necessary to fit a calibrated
PRT/Current option board so that step 8 can be completed

(this calibrates for the initial accuracy of the on-board voltage reference).