SIMULATE ACCURACY (0.1 deg C/F resolution)

THERMOCOUPLE

TYPE

TEMPERATURE

RANGE deg C

ACCURACY

Deg C

J

-210 to 150

150 to 1200

0.15

0.3

K

-270 to 190

190 to 1250

0.5

0.4

T

-200 to 150

150 to 400

0.4

0.5

R

-50 to 800

800 to 1750

0.8

2.0

S

-50 to 850

850 to 1750

0.9

2.0

B

100 to 1200

1200 to 1800

2.0

3.0

N

-270 to 260

260 to 1300

0.5

1.0

E

-50 to 1000

0.3

An additional correction representing the equivalent 1µV should be allowed for stray
thermal emf effects.

Millivolt Measure 0 to +/- 30mV

Resolution:

10µVAccuracy:

± 0.05% of f.s. ± 1 digit

Input resistance

100K Ohms

Milliamp Measure 0 to +/- 60mA

Resolution

20µA

Accuracy

± 0.05% of f.s. ± 1 digit

Input resistance

0.5 ohms