Time Electronics 1090 Portable Process Calibrator User Manual

Page 16

Advertising
background image

16

1090 Technical Manual

P a g e

| 16

Millivolt Source 0 to +/- 80mV

Accuracy

0.02% of f.s. (16uV)

Resolution

5uV

Output resistance

10 ohm

Note: For output settings below 8mV an increased resolution of 0.5uV is automatically
switched in. The accuracy for this part of the span is also increased to +/- 4uV

Milliamp Source 0 to 80mA

Accuracy

0.02% of f.s. (16uA)

Resolution

5uA

Maximum load (24V drive)

300R/80mA, 480R/50mA, 1.2K/20mA

Inching

Three levels of increment, 0.1, 1 or 10 for degC/degF, and 1, 10 or 100 uV/uA for voltage
current. The lowest of these represents the highest setting resolution and provides the most
precise control of the output.

24V process loop drive mode

A process loop can be driven at 24V up to 60mA by selecting the Milliamp source mode and
setting it at 60mA (or a lower level if required).

PT100 SIMULATION

14 set temperatures. Specification within ± 0.1% of DIN 43760 resistance values on all
settings.

Reading deg C

-100.0

-50.0

-20.0

0.0

20.0

50.0

100.0

200.0

Accuracy deg C

0.3

0.3

0.3

0.3

0.3

0.3

0.3

0.5

Reading deg C

300.0

400.0

500.0

600.0

700.0

800.0

Accuracy deg C

0.5

0.5

0.5

0.7

0.7

0.7

PT100 MEASURE (0.2 degC or degF resolution)

0 to 300.0 ohms, 2 wire.

Reading deg C

-100.0

-50.0

0.0

100.0

150.0

200.0

250.0

300.0

Accuracy deg C

0.8

0.8

0.7

0.7

0.7

0.7

1.0

1.0