Confi g Function Tab

8

CrystalControl Operation Manual

Modules: Temperature Module (RTD100)

The following features are unique to the Temperature Modules (RTD100):

Temperature Coeffi

cient of Resistance (TCR):

Select your appropriate platinum RTD Temperature Coeffi

cient of Resistance (TCR) from the pull down menu.

When selected, the pre-populated Callendar-Van Dusen Coeffi

cients will appear. See the nVision Operation Manual for further details on RTDs and the

RTD100 module.

Base Resistance (Ro):

Adjust your RTD resistance at 0°C. Factory default is 100Ω.

Lead Type:

Adjust your RTD sensor wiring between 2-, 3-, or 4-wire methods.

Typical view of RTD100 Module features in CrystalControl.