Kaman digiVIT User Manual

Page 12

Advertising
background image

www.kamansensors.com

PART NO: 860525-001

Last Revised 9/16/2013

- 12 -

4.8.3 Clear Zero

This will clear the zero offset and return to the calibrated absolute output.

4.8.4 3 Pt. Adjust

The 3 point adjust makes a slight adjustment to an existing calibration. It allows for correction of
installation or slight loading errors in gain and offset. For this calibration, the sensor must be
positioned at MIN, MID, and MAX with data taken at each point. This adjustment does not
optimize the output further; it simply corrects scale, offset, and small linearity errors.
Temperature coefficients are not changed. It will remove any offset from zeroing the sensor
output. This option is used when more accuracy, but not a full calibration is required.