Probe specifications – Kaman digiVIT User Manual

Page 25

Advertising
background image

www.kamansensors.com

PART NO: 860525-001

Last Revised 9/16/2013

- 25 -

Characterized with an aluminum target

2U

4U

9U

12U

16U

26U

38U

51U

Offset

inch

0.002

0.005

0.010

0.016

0.020

0.032

0.050

0.100

(mm)

(0.05)

(0.13)

(0.25)

(0.40)

(0.50)

(.80)

(1.20)

(2.50)

Short Range

inch

0.010

0.025

0.050

0.080

0.100

0.160

0.250

0.300

(mm)

(0.25)

(0.60)

(1.25)

(2.00)

(2.50)

(4.00)

(6.00)

(7.5)

Standard Range

inch

0.020

0.050

0.100

0.160

0.200

0.320

0.500

0.600

(mm)

('0.50)

(1.30)

(2.50)

(4.00)

(5.00)

(8.00)

(12.0)

(15.0)

Extended Range

inch

0.030

0.070

0.150

0.240

0.320

0.500

0.800

1.000

(mm)

(0.75)

(1.75)

(3.75)

(6.00)

(8.00)

(12.5)

(20.0)

(25.0)

Typical Specifications for stardard range, aluminum target, standard cable length

Nonlinearity 6 pt calibration

±%FS

<0.3%

Nonlinearity 21 pt calibration

±%FS

<0.1%

Static Resolution

RMS%FS

<0.001%

Resolution @ 100 Hz

RMS%FS

<0.01%

Resolution @ 1000 Hz

RMS%FS

<0.02%

Thermal Sensitity

%FS°F

0.05

(%FS°C)

(0.1)

Thermal Sensitity w/temp

%FS°F

0.01

comp calibration

(%FS°C)

(0.02)

Probe Specifications