Kaman digiVIT User Manual

Page 23

Advertising
background image

www.kamansensors.com

PART NO: 860525-001

Last Revised 9/16/2013

- 23 -

Command Passlevel

Required

Parameters

Returns Comments

System

RXR

NONE

NONE

Revision

Read Firmware Revision

U

NONE

Sets Passlevel

Passlevel 0-2

Unlocks for higher level control

U KAMAN<ret> -- unlocks to User Level

U<ret> -- sets password to normal level

MISC TIMER

WC

NONE

NONE

Seconds

Watch Clock (seconds since power on)

WT

NONE

NONE

Ticks

Watch Ticks (Ticks since power on)

CALIBRATION

CS

NONE

NONE

--

Saves the calibration and all parameters

CF NONE

NONE

Frequency

Set Optimal Full Scale – Sets Frequency and

Optimal Coil Current – sensor must be positioned

at max displacement from sensor. This is the first

thing done in a calibration.

CZ NONE

NONE

--

Sets Zero and sets gains and offsets (Executed

after CF) – sensor must be positioned at the

minimum displacement from sensor. This is the

second thing done in a calibration

CP x

NONE

X = 0-20

AD Reading

Cal Point (0-20) for piecewise linearization –

sensor must be positioned at the correct percent of
full scale – i.e. 0=0%, 1=5%,2 =10% … 20=100% -

- direction (min to max or max to min) does not

matter as long as it changes monotonically. For 6
point polynomial cals only use 0-5 points and they

must be spaced 20% apart ... i.e. pt 0 is 20%, pt1

is 40% ... pt 5 = 100% of range. For a 2 point cal

pt 0 = 0% pt 1 = 100%.

CD NONE

NONE

Status

Completes the 21 point calibration and sets up the

tables for operation

C6 NONE

NONE

Status

Completes the 6 point polynomial cal

C2

NONE

NONE

Status

Completes the 2 point linear cal

CAZ

NONE

NONE

--

Cal Adjust zero point

CAM

NONE

NONE

--

Cal Adjust mid point

CAF

NONE

NONE

--

Cal Adjust full scale point

CAD

NONE

NONE

Status

Complete 2 Point Adjust

CT 0

NONE

NONE

--

Collect Temp Comp Data for D1T1

CT 1

NONE

NONE

--

Collect Temp Comp Data for D1T2

CT 2

NONE

NONE

--

Collect Temp Comp Data for D2T1

CT 3

NONE

NONE

--

Collect Temp Comp Data for D2T2

CT 4

NONE

NONE

Status

Completes Temperature Compensation for Sensor

CT 5

NONE

NONE

Status

Completes Temperature Compensation for

Electronics

Advertising