APPENDIX A: COMMAND LIST

The following are the low level commands for the user to program an interface. Many of the
setup commands are read or write. The command starts with an 'S' for a write (such as SCLL
20000) or an 'R' for a read (RCLL will read back the setting). Settings are stored in non-volatile
system memory.

Command Passlevel

Required

Parameters

Returns Comments

Monitor Commands

MD NONE

NONE

Distance

Output

Monitor Distance Output 100000=100%

ML

NONE

NONE

Monitor Limit

Monitor Limit 0= in range 1=low 2 = hi

MR

NONE

NONE

Monitor Relay

Monitor Relay 0= Open 1 = Closed

AutoZero Function

ZZ

NONE

NONE

--

Zeroes the output

ZC

NONE

NONE

--

Clears the zero on the output

Limit Setup

S(R)CLP x

USER

X = polarity

Polarity

SET(READ) Limit Polarity 0 = NO 1= NC

S(R)CLH xxxx

USER

xx = Hi Limit

Hi Limit

SET(READ) Limit Hi (Signed Long Int)

S(R)CLL xxxxx

USER

xx = Lo Limit

Lo Limit

SET(READ) Limit LO (Signed Long Int)

S(R)CLD xxxx

USER

xx = Hysteresis

Hysteresis

SET(READ) Limit Deadband (Hysteresis)

(Signed Long Int)

Ethernet Setup

S(R)EA xxx xxx xxx xxx

USER

xx xxx xxx xxx =

IP Address

IP Address

SET(READ) Ethernet IP Address

Note: Separate Octets with spaces not '.'

S(R)EW xxxxx

USER

xx = port

UDP Writer Port

SET(READ) Ethernet Writer Port

S(R)ER xxxxx

USER

xx = port

UDP Reader

SET(READ) Ethernet Reader Port

REM

NONE

MAC

READ Ethernet MAC Address

Output Setup

S(R)UI x

USER

x = setting

Setting

SET(READ) User Current Output

0=0-20mA
1=4-20mA

S(R)UV x

USER

x = setting

Setting

SET(READ) User Voltage output

0=0-5V

1 =+/-5V
2=0-10V

3=+/-10V