UEi Test Instruments KM940 User Manual

Page 20

Advertising
background image

- 20 -

5.2.4 Display Menu

L

IGHT

MODE
CONTRAST

: OFF
:
8-PAGE
:
DEFAULT


Allows the configuration of the display to be changed.

LIGHT :

Choose from ON or OFF.


MODE :

Select 4 or 8 Page Mode or Line Scroll Mode as detailed in section 4.3 Main
Displays.


CONTRAST :The contrast is set to a DEFAULT value or can be adjusted

↑ LIGHTER or ↓ DARKER. Use the and keys to

adjust.

5.2.5. Set-Up Menu


The set up menu allows the following parameters to be set / altered.

• Language.
• Automatic calibration time
• CO gas alarm
• NOx percentage for calculation
• Date and time
• Printout Header

L

ANG

CO AL

CALEN

: ENGLISH
ARM : 400
DAR

ZERO : 3
NOx% : 5
HEADER


Parameter Description

Settings

LANG :

Changes the analysers displayed and printed
language.

ENGLISH
SPANISH
DUTCH
FRENCH
ITALIAN

ZERO :

Allows setting of the Autocalibration time in
minutes. Care must be taken when changing this
parameter as sensors may drift from zero if too
short a time is used. Kane International advise 3
minute countdown.

2-6 minutes

CO ALARM :

Allows an alarm level to be set on for the CO
reading. This is set as a default at 1000 ppm.

OFF
0-4000 ppm