UEi Test Instruments KM940 User Manual

Page 9

Advertising
background image

- 9 -

4.

NORMAL START UP SEQUENCE

4.1 Every Time You Use The Analyser

BEFORE SWITCH-ON CHECK THAT:


the particle filter is not dirty


the water trap and probe line are empty of water


all hose connections, etc, are properly made


the probe is sampling CLEAN AMBIENT air


the water trap is correctly fitted and the instrument upright


the flue temperature is connected


Switch ON the instrument by pressing

4.2 Automatic

Calibration


During this sequence the analyser pumps fresh air into the sensors to allow toxic sensors (if
fitted) to be set to zero and the Oxygen sensor to be set to 20.9 %.

After switch-on the analyser will briefly display header information :-

Kane International

(44)-1707-375550


And then show the countdown screen :-

ZERO CAL
Time : 180

FRESH AIR PURGE


The calibration time will count down in seconds to zero. Calibration time may be changed
from 2 to 6 minutes. See Set-Up menu section 5.2.5.

Note ! Three minutes is recommended to allow the sensors to stabilise fully. Anything less

than this may result in drift of the toxic and oxygen sensors in clean ambient air.


To obtain the quoted specification an instrument should be calibrated with clean ambient air
at standard temperature and pressure (STP).