Document No.001132, Revision 2.0

NodeID Livelink : 14140190

Agfa HealthCare

14 June, 2007

Attribute Name

Tag

Value

Execution State Info

(2100, 0030)

Printer dependent.

Print Priority

(2000, 0020)

HIGH, MED, LOW

Creation Date

(2100, 0040)

Date of print job creation.

Creation Time

(2100, 0050)

Time of print job creation

Printer Name

(2110, 0030)

User defined name identifying the printer

Originator

(2100, 0070)

DICOM AE title that issued the print operation

2.2.2.3.1.5

SOP Specific Conformance Presentation LUT SOP Class

(1.2.840.10008.5.1.1.23)

The Presentation LUT Information Object is an abstraction of a Presentation LUT. The objective of the
Presentation LUT is to realize image display tailored for specific modalities, applications, and user
preferences. It is used to prepare image pixel data for display on devices that conform to the Grayscale
Standard Display Function.
The output of the Presentation LUT is Presentation Values (P-Values).

A N-CREATE is issued by "Workstation" (SCU) to create a Presentation LUT SOP Instance. The supported
Presentation LUT attributes are listed below:

Table 2.2-35: Supported Attributes for Presentation LUT

Tag

Name

Supported

Default

(2050,0020)

Presentation LUT Shape

IDENTITY (Note)

IDENTITY

Note:

‘Presentation LUT Sequence’ is not supported.

"Workstation" (SCU) will process the N-CREATE confirmation and response Status codes. The status codes
listed below are recognized:

Table 2.2-36: N-CREATE confirmation and response Status codes

Code

Status

Meaning

0000

Success

Presentation LUT successfully created

B605

Warning

Requested Min or Max Density outside of printer’s operating range. The printer will use its
respective minimum or maximum value instead.