Note:

1.

The requirement for the final state is that which applies at the time that the Performed Procedure Step
Status (0040,0252) is N-SET to a value of COMPLETED or DISCONTINUED. It is only described if it is
different from the SCP requirement for the N-CREATE.

2.

The Performed Series Sequence (0040,0340) may not be empty (zero length) at the time that the
Performed Procedure Step Status (0040,0252) is N-SET to a value of COMPLETED or
DISCONTINUED. In other words a Series must exist for every Performed Procedure Step, though it
may contain no Images or Non-Image Composite objects, if none were created.

3.

Only attributes that are specified in a SOP Instance at N-CREATE may later be updated through the N-
SET. If "Workstation" wishes to use the PPS Discontinuation Reason Code Sequence (0040, 0281), it
creates that attribute (zero-length) during MPPS N-CREATE.