Media storage application profile, Options – AGFA HE/001132 User Manual

Page 69

Advertising
background image

HE/001132

Page 69 of 92

Document No.001132, Revision 2.0

NodeID Livelink : 14140190

Agfa HealthCare

14 June, 2007

Exceptions exist for attributes which are necessary to correctly import or interpret the image. Therefore the
following values are kept or replaced by generated values:

Table 3.2-2: Necessary Values when performing an Anonymous Export.

Attribute

Tag

Set to …

Patient’s Name

(0010,0010)

“Anonymous”

Study Instance UID

(0020,000D)

Keep value

Modality

(0008,0060)

Keep value

Series Instance UID

(0020,000E)

Keep value

Laterality

(0020,0060)

Keep value

Series description

(0008,103E)

Keep value

Protocol name

(0018,1030)

Keep value

Body Part Examined

(0018,0015)

Keep value

Smallest Pixel Value in Series

(0028,0108)

Keep value

Largest Pixel Value in Series

(0028,0109)

Keep value

Presentation Intent Type

(0008,0068)

Keep value

SOP Instance UID

(0008,0018)

Keep value

SOP Class UID

(0008,0016)

Keep value

Specific character set

(0008,0005)

Keep value

Patient Orientation

(0020,0020)

Keep value

Image type

(0008,0008)

Keep value

Image comments

(0020,4000)

Keep value

Presentation LUT shape

(2050,0020)

Keep value

3.2.1.1.1.1

Media Storage Application Profile

The Offline-Media Application Entity supports the STD-GEN-CD Application Profile.

3.2.1.1.1.1.1

Options

The Offline-Media Application Entity supports the SOP Classes and Transfer Syntaxes listed in the Table
below:

Table 3.2-3: IOD’S, SOP Classes and Transfer syntaxes for offline media

IOD

SOP Class UID

Transfer Syntax

Transfer Syntax UID

CR Image Storage

1.2.840.10008.5.1.4.1.1.1

Explicit VR Little Endian

1.2.840.10008.1.2.1

Digital X-Ray Image

Storage – For Presentation

1.2.840.10008.5.1.4.1.1.1.1

Explicit VR Little Endian

1.2.840.10008.1.2.1