Getting Started

Getting Started

Getting Started

The instructions in this chapter show you how to install your Agilent 83438A.
You should be able to finish these procedures in about ten to twenty minutes.
After you’ve completed this chapter, continue with Chapter 2, “Making Mea-
surements”
.

Refer to Chapter 3, “Specifications and Regulatory Information” for informa-
tion on operating conditions such as temperature.

If you should ever need to clean the cabinet, use a damp cloth only.

W A R N I N G

This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact.
Any interruption of the protective conductor inside or outside of the
product is likely to make the product dangerous. Intentional
interruption is prohibited.

C A U T I O N

This product has autoranging line voltage input. Be sure the supply voltage is
within the specified range.

C A U T I O N

VENTILATION REQUIREMENTS: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4

°C for every 100 watts dissipated in the

cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.

C A U T I O N

This product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2, per IEC 1010 and 664 respectively.