Specifications and Regulatory Information

Specifications and Regulatory Information

Specifications and Regulatory Information

This chapter lists specification and characteristics of the instrument. The dis-
tinction between these terms is described as follows:

• Specifications describe warranted performance over the temperature range

0

°C to +45°C and relative humidity <95% (unless otherwise noted). All speci-

fications apply after the instrument’s temperature has been stabilized after 15
minutes of continuous operation.

Characteristics provide useful information by giving functional, but nonwar-

ranted, performance parameters. Characteristics are printed in italics.

Calibration cycle
This instrument requires periodic verification of performance. The instrument
should have a complete verification of specifications at least once every two
years.

Instrument block diagram
You can find a block diagram of the instrument in the front of this book. Refer
to “Block Diagram of Agilent 83438A” on page ii